X-ray single crystal diffractometer Rigaku Synergy S

X-ray single crystal diffractometer Rigaku Synergy S
  • Skraćeni naziv instrumenta

    Single crystal X-ray diffractometer

    Inventarni broj

    49050

    Opći opis

    Rigaku XtaLAB Synergy-S is an X-ray diffractometer for X-ray diffraction on a single crystal. It has two PhotonJet-S microfocus sources of X-rays, with anodes made of molybdenum (Mo) and copper (Cu). These third-generation X-ray tubes in combination with microfocus optics represent strong sources of collimated radiation of wavelengths of 0.071 and 0.154 nm. XtaLAB Synergy-S monocrystalline X-ray diffractometer comes with a four-circuit goniometer, in the so-called. kappa configuration, and has a high speed motor for crystal orientation. The system is equipped with a highly sensitive and fast X-ray detector (HyPix-6000HE).
    Very small samples of crystals of small organic molecules and crystals of biological macromolecules such as proteins, less than 10 micrometers in size, can also be measured on this instrument. The diffractometer is equipped with a cryostat that allows the collection of data on samples in the temperature range from 90 to 400 K.

    Uža područja primjene

    crystallography, condensed matter physics, biochemistry

    Namjena instrumenta

    crystal structure determination from organic, inorganic and biological samples

    Popis usluga

    determination of the crystal structure of small molecules; determination of crystal structure of biological macromolecules; determination of charge density; determination of crystal structure under high pressure;

    Opis na stranicama zavoda/laboratorija

    https://www.irb.hr/eng/Research/Capital-equipment/Rendgenski-difraktometar-za-monokristale-Oxford-Diffraction-Xcalibur-Nova-R

    Tijelo koje je financiralo nabavku opreme

    ,

    Project title

    Open scientific infrastructure platforms for innovative applications in the economy and society (O-ZIP)

    Type of instrument

    ,

    Type of the analysis

    , , ,

    Applications

    , , , , , , , , , , ,

    Ključne riječi

    crystallography, protein crystallography, molecule structure, charge density

    Procijenjeni broj korisnika

    50

    Year of manufacture

    2021

  • Ustanova en

    Institut Ruđer Bošković
    Bijenička cesta 54, Zagreb

    Zavod

    Department of physical chemistry

    Laboratorij

    Laboratory for chemical and biological crystallography

    Grad

    Zagreb

    Ulica i broj

    Bijenička 54

    Krilo/Kat/Soba

    3. wing, 109

  • Način korištenja instrumenta

    , ,

    Cjenik korištenja

    The price per sample is 500kn for users from the RBI and 720kn for external users.

    Popis ovlaštenih korisnika

    Zoran Štefanić
    Krešimir Molčanov
    Aleksandar Višnjevac
    Valentina Milašinović
    Petra Stanić

    Broj istovrsnih instrumenata u ustanovi

    1

  • Ime i prezime

    Zoran Štefanić

    Titula

    Dr.

    E-mail

    zoran.stefanic@irb.hr

    Telefon

    01 4571405

    Mobitel

    0912804972

  • Ime i prezime

    Krešimir Molčanov

    Titula

    Dr.

    E-mail

    kresimir.molcanov@irb.hr

    Telefon

    4571405

  • Model

    Rigaku XtaLAB Synergy-S

    Mrežna (URL) adresa instrumenta na stranicama proizvođača

    https://www.rigaku.com/products/crystallography/synergys

    Princip rada i mjerna tehnika

    An X-ray diffractometer for single crystals illuminates the crystal sample with a beam of X-rays of a fixed wavelength. The structure of the molecules that make up the crystal is determined from a series of diffraction images obtained in various positions of the crystal.
    It is possible to determine the positions of atoms in molecules at the atomic level. The instrument can record data on molecules from just a few atoms to large molecules with tens of thousands of atoms such as proteins.

    Radno i mjerno područje

    positions of atoms on a subnanometar scale, temperature range 90-400K

    Proizvođač

    Rigaku

    Mrežna (URL) adresa proizvođača

    https://rigaku.com

    Purchase price

    2.8

    Estimated current value of equipment

    2.8

    Vijek trajanja

    up to 20 years

    Ostale tehničke karakteristike

    • Four-circle kappa goniometer for high precision crystal positioning and orientation (sphere of confusion less than 10 μm)
    • Equiped with dual wavelength X-ray source with Cu and Mo anodes mounted on the same goniometer
    • Two dimensional, low noise, direct photon counting X-ray detector with fast readout times (in ms) and high dynamic range (at least 24 bits) and no dimension reduction (with eg. optical tapers)
    • Exchangable collimators with apertures 0.2-0.7 mm
    • Safety cabinet with wide opening for easy access to the goniometer, with internal lighting with low heat production
    • Closed circuit, reliable and robust cooling of X-ray tube and detector (air or water)
    • High voltage generator power supply with high voltage and current stability (50 kV, 1mA)
    • Two goniometer heads with possibility of mounting the crystals on glass needles and in cryo loops
    • High intensity microfocs X-ray tubes with focusing of X-rays onto sample crystal and producing low divergence X-ray beam.
    • Liquid nitrogen cryo device capable of maintaining the sample at 100 K with no interruptions. One movable liquid nitrogen vessel of at least 130 L capacity for refilling the system.
    • PC for operating the instrument with software for data acquisition on small molecule and protein samples installed and free upgrades.

    Mrežna (URL) adresa s tehničkim karakteristikama

    https://www.rigaku.com/products/crystallography/synergys

    Popratna i dodatna oprema

    Oxford Cryostream 800
    High Pressure Kit

  • Slike en
    Rigaku XtaLAB Synergy S
    Rigaku XtaLAB Synergy S
    Detector and goniometer
    Detector and goniometer
    Collimators and optics
    Collimators and optics

Natrag en